IVD Manufacturer Data
- The most knowledgeable source of information on instrument performance and likely effects of possible interferences with an HBOC is the IVD manufacturer of your laboratory instrument. The manufacturer should be contacted via its Technical Service Department.
- Some IVD instrument manufacturers have presented or published results from investigations of possible interferences with Hemopure, Oxyglobin or both. The following IVD manufacturers are especially likely to have interference data:
|
IVD equipment manufacturer |
Instrument model |
Equipment type |
|
Bayer HealthCare Diagnostics Division Tarrytown, NY |
Advia 120 |
Hematology |
|
Brea, CA |
Synchron CX7 |
Chemistry |
|
Newark, DE |
Dimension RxL Max
Dimension Xpand |
Immunochemistry & Chemisty; Chemistry |
|
Lexington, MA |
GEM Premier 3000 Synthesis |
Point-of-care; Critical Care |
|
East Windsor, NJ |
Point-of-care |
|
|
Rochester, NY |
VITROS Microslide, VITROS ECi Immunodiagnostic, VITROS MicroTip assay |
Chemistry, Immunodiagnostic, Chemistry |
|
Tutzing, Germany |
Cobas Integra 700 Hitachi 917 |
Chemistry |
Hemopure Interference Testing Performed by IVD Manufacturer or Contract Laboratory
i-STAT System® Assay Limitations with Hemopure
i-STAT System Whole Blood Interference Limits for Hemopure.#
Hemopure interferences on whole blood measurements made on the i-STAT System have been investigated using CG8+, EC8+, CG4+ and CREA cartridges to determine the maximum whole blood concentration of Hemopure (g/dL) that can be tolerated for individual assays. Assay results should be comparable across all relevant i-STAT cartridges types.
|
Analyte |
Maximal Allowable Hemopure Concentration (g/dL) |
|
Blood urea nitrogen (BUN) |
6.5 |
|
Calcium, ionized |
6.5 |
|
Chloride |
6.5 |
|
Creatinine |
3.9 |
|
Glucose |
6.5 |
|
Hematocrit |
5.2 |
|
Calculated Hemoglobin[*] |
0 |
|
pCO2 |
6.5 |
|
pH |
6.5 |
|
pO2 |
6.5 |
|
Lactate |
5.2 |
|
Potassium |
6.5 |
|
Sodium |
6.5 |
#This report reflects a joint study commissioned by OPK Biotech and executed with the support of i-STAT (Abbott Point of Care). This should not be perceived as a marketing claim or a claim regarding performance of Abbott Point of Care.
[*]Hemoglobin, in g dL-1, is calculated by multiplying the measured hematocrit value, in %PCV, by 0.34.
Dade Behring HBOC-201 Interference Results
Interferences have been tested on the Dade Behring Dimension® Xpand® and Dade Behring Dimension® RxL Max® systems spiked with HBOC-201 concentrations ranging between 0 and 6.4 g/dL.
When the Dade Behring Dimension® Xpand® and Dimension® RxL Max® are operated using Dade Behring products and according to the manufacturer's instructions, the following results were observed. Since HBOC-201 concentrations were tested up to an upper limit of 6.4 g/dL, patient samples with a Hemopure concentration above this concentration should not be analyzed for these assays due to unknown interference.
Dimension® Xpand® and Dimension RxL Max® Systems
|
Analyte |
HBOC-201 (g/dL) above which assay results are inaccurate or interference unknown |
Comments |
|
Na |
ok to 6.4 |
NA results decrease by about 1 mmol/L per 1 g/dL HBOC-201 |
|
K |
ok to 6.4 |
K results decrease by about 0.03 mmol/L per 1 g/dL HBOC-201 |
|
Cl |
ok to 3.3 |
CL results decrease by about 1.5 mmol/L per 1 g/dL HBOC-201 |
|
ECO2 |
ok to 0.1 |
ECO2 results decrease, Absorbance error over 3 g/dL HBOC-201 |
|
BUN |
ok to 6.4 |
BUN results increase slightly by about 0.2 mg/dL per 1 g/dL HBOC-201 |
|
CREA |
ok to 0.3 at 2.0 mg/dL CREA or 0.8 at 6.0 mg/dL CREA |
CREA results decrease by 0.7 mg/dL per 1 g/dL HBOC-201 |
|
GLU |
ok to 0.8 at 75 mg/dL GLU or 2.5 at 240 mg/dL GLU |
GLU results decrease by about 10 mg/dL per 1 g/dL HBOC-201 |
|
CA |
ok to 2.0 |
CA results increase by 0.3 mg/dL per 1 g/dL HBOC-201 |
|
PHOS |
ok to 2.0 |
PHOS results increase, and may have Abnormal Assay error code above 3 g/dL of HBOC-201 |
|
MG |
ok to 6.5 |
MG results stable within the larger of 0.2 mg/dL or 10% |
|
TP |
ok to 0.5 |
TP results Increased about 0.6 g/dL per 1 g/dL HBOC-201 |
|
ALB |
ok to 2.5 |
ALB results lowered about 0.7 g/dL at 6.5 g/dL HBOC-201 |
|
URCA |
ok to 0.5 |
URCA results with HBOC-201 over 1 g/dL have absorbance error flag |
|
TBIL |
ok to 0.4 |
TBIL results with HBOC-201 over 0.5 g/dL have hemoglobin error flag |
|
DBIL |
ok to <0.1 |
DBIL result decreases about 0.2mg/dL for each 1 g/dL of HBOC-201, have hemoglobin error over 0.5 g/dL HBOC-201 |
|
CHOL |
ok to 1.3 at 100 mg/dL CHOL or 2.0 at 200 mg/dL CHOL |
CHOL result decreases about 8 mg/dL for each 1 g/dL of HBOC-201 |
|
AHDL |
ok to 0.7 |
AHDL result decreases about 12% for each 1 g/dL of HBOC-201 |
|
TGL |
ok to 0.2 at 50 mg/dL TGL or 0.7 at 130 mg/dL TGL |
TGL result increases about 20 mg/dL for each 1 g/dL of HBOC-201 |
|
AST |
ok to 0.5 |
Error code with erroneous results |
|
ALT |
ok to 0.5 |
Error code with erroneous results |
|
ALP |
ok to 1.0 |
Error code with erroneous results |
|
AMY |
ok to 1.0 |
Error code with most erroneous results |
|
CK |
ok to 0.1 g/dL at 100 U/L CK or 0.5 g/dL at 350 U/ L CK |
HBOC-201 over 1 g/dL substantially increases CK results |
|
LDH |
ok to 2.0 |
HBOC-201 over 2 g/dL decreases LDH results |
|
GGT |
ok to 0.5 |
Error code with erroneous results |
|
MMB |
ok to 5.0 |
- |
|
MYO |
ok to 6.5 |
- |
|
CTNI |
ok to 4.0 |
CTNI results increase about 2% per 1 g/dL HBOC-201 |